Catalog Number 762165 |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that 20 paxgene® blood rna tubes cracked after use, exposing the staff to blood, though they reportedly wore ppe and were not sent for testing.This complaint was created to capture 1 of 2 related incidents.The following information was provided by the initial reporter: "stated she is experiencing cracking of tubes after use, workers were exposed to blood but no testing was done." "worker was not sent for testing.Ppe was exposed to blood.".
|
|
Event Description
|
It was reported that 20 paxgene® blood rna tubes cracked after use, exposing the staff to blood, though they reportedly wore ppe and were not sent for testing.This complaint was created to capture 1 of 2 related incidents.The following information was provided by the initial reporter: "stated she is experiencing cracking of tubes after use, workers were exposed to blood but no testing was done." "worker was not sent for testing.Ppe was exposed to blood.".
|
|
Manufacturer Narrative
|
Investigation: investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: as there was no sample or photo available for evaluation, a root cause could not be determined.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|
|
|