A field service engineer (fse) was at the customer's site to address the reported issue.The fse confirmed the issue on the error log and was able to reproduce the error by performing wash prime and noted b/f (bound free) probe 1 did not evacuate the wash well after purge.The fse removed b/f probe 1 and 2 and performed prime wash and verified that b/f probe 1 was not evacuating liquid after purge.The fse switched the fluid fittings from b/f probe 1 to b/f probe 2 with the 2-way solenoid valves and noted the fault moved to b/f probe 2; the fse replaced the 2-way solenoid valve.The fse performed several wash primes and completed daily check with no issues or errors.Analyzer validation was performed through completing quality control (qc).Qc results passed within the manufacturer's published specifications.The aia-2000 analyzer conformed to the manufacturer's performance specifications and returned to operation.No further action required by field service.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4), from 10mar2018 to aware date 10apr2019.One other similar complaint was identified during the search period.The aia-2000 operator's manual, section appendix 4: error messages, states the following: [2235] b/f probe 1 suction failure.Cause: the overflow sensor detected liquid after washer suction action was completed.The measurement result will be flagged (wu flag).Solution : clean b/f probe 1.(see chapter 9, "maintenance") if retry fails, contact tosoh service center or local representatives.The most probable cause of the reported issue is attributed to the faulty 2-way solenoid valve.
|
A customer reported getting system error 2235, b/f (bound free) probe 1 suction failure on the aia-2000 analyzer.The customer replaced the tip on the b/f probe, but the error persisted.A field service engineer (fse) was dispatched to address the reported issue, which resulted in delayed reporting of intact parathyroid hormone (ipth) patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|