• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TALAR COMP,SINGLE COATED US VERS X-SMALL, LEFT; PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TALAR COMP,SINGLE COATED US VERS X-SMALL, LEFT; PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED Back to Search Results
Catalog Number 400253
Device Problem Device Slipped (1584)
Patient Problem Failure of Implant (1924)
Event Date 12/20/2018
Event Type  Injury  
Manufacturer Narrative
Device was not returned.If additional information becomes available, it will be provided in a supplemental report.Still implanted.
 
Event Description
It was reported from the clinical team that the subject received a star in the left ankle on (b)(6) 2017.On (b)(6) 2018, subject presented with left talar subsidence with subtalar cystic change resulting in device subsidence.The clinical investigator deems this ae as severe and related to the device.The subject is expected to receive surgical intervention for her left ankle in the coming months.
 
Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device and medical records must be available in order to determine the root cause of the complaint event.The device inspection was not possible as the product was not returned for investigation.Cysts (osteolysis), pain and implant dislocations are listed as adverse effects in the ifu; furthermore, a warning regarding activities like running and jumping as a potential cause for implant failures.In addition, patient has a bmi of 43.9 and is obese.The misalignment was most likely caused by the cysts.A correlation between the implants and the cyst cannot be excluded.Cysts and implant dislocations are classified as known adverse effect.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported from the clinical team that the subject received a star in the left ankle on (b)(6) 2017.On (b)(6) 2018, subject presented with left talar subsidence with subtalar cystic change resulting in device subsidence.The clinical investigator deems this ae as severe and related to the device.The subject is expected to receive surgical intervention for her left ankle in the coming months.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TALAR COMP,SINGLE COATED US VERS X-SMALL, LEFT
Type of Device
PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key8579808
MDR Text Key144093947
Report Number0008031020-2019-00390
Device Sequence Number1
Product Code NTG
Combination Product (y/n)N
PMA/PMN Number
P050050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/30/2019
Device Catalogue Number400253
Device Lot Number1428093
Was Device Available for Evaluation? No
Date Manufacturer Received05/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient Weight102
-
-