• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0535FC
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/09/2019
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.An error message u509 ultrasonic was displayed when a probe check was performed.A hairline crack on the probe unit was observed.Based on visual inspection on the device there were some tissue build up.The teflon pad was inspected and found it has normal wear, no metal exposed, no abnormality.Based on similar reported events, the most probable cause for the crack on the probe is attributed to the probe coming into contact with a hard or metallic surface during activation.The instruction manual contains several warning statements in an effort to prevent damage to the probe.¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.¿.
 
Event Description
Olympus was informed that during a total laparoscopic hysterectomy procedure probe damage error occurred.Procedure was completed using a similar device.Patient experienced minimal bleeding but did not require medical or surgical intervention.The device was inspected prior to use and no abnormalities were observed.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP
Type of Device
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
connie tubera
2400 ringwood avenue
san jose, CA 95131
4089355124
MDR Report Key8580519
MDR Text Key144138347
Report Number2951238-2019-00804
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04953170337574
UDI-Public04953170337574
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 05/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTB-0535FC
Device Lot NumberKR815800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-