• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF283
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2019
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the balloon catheter, 2af283 with lot number 95559, was returned and analyzed.Visual inspection of the balloon catheter showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used 29 application.System notice (b)(4) ¿the safety system has detected fluid in the catheter and stopped the injection¿ was triggered when catheter connected to the console.Dissection / pressure testing showed a double balloon breach.In conclusion, the balloon catheter failed the returned product inspection due to the double balloon breach.If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the balloon catheter subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8581004
MDR Text Key144161436
Report Number3002648230-2019-00324
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/30/2020
Device Model Number2AF283
Device Catalogue Number2AF283
Device Lot Number95559
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2019
Date Manufacturer Received04/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
-
-