COOK INC THREE-WAY PLASTIC STOPCOCK; KGZ ACCESSORIES, CATHETER Back to Search Results
Catalog Number PTWS-2FLL-MLL-R
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/29/2019
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) number = exempt.This report includes information known at this time.A follow-up report will be submitted at the conclusion of the investigation or, when additional relevant information becomes available.
 
Event Description
It was reported, a patient of unknown gender or age was undergoing a transarterial chemoembolization (tace) procedure using a three-way plastic stopcock.The stopcock was attached and the physician was infusing the drug through the stopcock when the drug leaked out of a crack in the device.A new three-way plastic stopcock was attached and the procedure was successfully completed without adverse patient effect.
 
Manufacturer Narrative
Investigation/evaluation: a review of the complaint history, device history record, quality control, visual and functional inspection of the returned device was conducted during the investigation.The visual inspection of the returned device confirmed a crack extending from the hex nut towards the center female connection.Functional testing confirmed the device leaked from the area of the crack.As no instructions for use are provided with this device, no review of instructions, indications, warnings, or precautions could be completed.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided and the examination of the returned product, investigation has concluded a cause for this event cannot be established.Appropriate measures have been initiated to address this failure mode.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE-WAY PLASTIC STOPCOCK
Type of Device
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8583117
MDR Text Key144177384
Report Number1820334-2019-01035
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier00827002002197
UDI-Public(01)00827002002197(17)240117(10)9454616
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/17/2024
Device Catalogue NumberPTWS-2FLL-MLL-R
Device Lot Number9454616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Date Manufacturer Received05/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-