LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).The field service representative replaced the motor end stage board hmf 0408 and subsequent functional verification testing found no further issues.The issue was solved and the device was returned to service.The replaced part was requested back to livanova (b)(4) for further investigation.The reported issue could be reproduced and traced back to a defective transistor.A review of the dhr did not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova (b)(4) received a report that rotating the pump rotor manually in the direction opposite to the operational one when a s5 roller pump was off, no alarm sounded.This was identified by a livanova field service representative during maintenance.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 89309
GM  89309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 89309
GM   89309
Manufacturer Contact
barbara galasso
14401 w. 65th way
arvada, CO 80004
2812287575
MDR Report Key8591290
MDR Text Key145975812
Report Number9611109-2019-00330
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse