• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problem Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 04/11/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom driver was not supporting a patient.The customer, a syncardia certified hospital, reported that the screw head on the freedom driver beat rate cover screw was stripped and could not be removed.
 
Manufacturer Narrative
Visual inspection confirmed the customer-reported issue of a damaged potentiometer cover torx screw.The root cause of the damaged potentiometer cover torx screw cannot be conclusively determined but is likely due to improper use (incorrect tool to remove or insert the screw).A specialized tool, the l torx wrench, is provided with the freedom driver toolkit to remove and screw in the potentiometer cover torx screw.F-900013-en, freedom driver system operator manual for us, describes instructions for adjusting the beat rate of the freedom driver in section 7.7.Labelling states, "unscrew the setting dial cover on the back of the driver by using the l torx wrench." if this tool was not used to insert or remove the screw, it is very likely that the torx screw would be damaged when adjusted with an improper tool.Despite the observed damage, the driver passed all sections of functional testing.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4744 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8593383
MDR Text Key144563269
Report Number3003761017-2019-00087
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2019
Date Manufacturer Received04/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-