Model Number 195274 |
Device Problem
Degraded (1153)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) discovered the issue while replacing the yellow level detector on the same hlm.Both level detectors have deteriorated pads and would not make proper contact with reservoir.Both level detectors were replaced.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.This complaint is related to (b)(4).
|
|
Event Description
|
It was reported that during use of the device for a non-clinical activity, the red level detector on the heart lung machine (hlm) was deteriorated.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed intermittent disconnect alert messages due to a deteriorated sensor membrane.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|