• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 195274
Device Problem Degraded (1153)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) discovered the issue while replacing the yellow level detector on the same hlm.Both level detectors have deteriorated pads and would not make proper contact with reservoir.Both level detectors were replaced.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.This complaint is related to (b)(4).
 
Event Description
It was reported that during use of the device for a non-clinical activity, the red level detector on the heart lung machine (hlm) was deteriorated.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed intermittent disconnect alert messages due to a deteriorated sensor membrane.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8594708
MDR Text Key144570783
Report Number1828100-2019-00222
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number195274
Device Catalogue Number195274
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Was the Report Sent to FDA? No
Date Manufacturer Received06/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-