• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY VDW GMBH EDDY IRRIGATION TIP; SYRINGE, PERIODONTIC, ENDODONTIC, IRRIGATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY VDW GMBH EDDY IRRIGATION TIP; SYRINGE, PERIODONTIC, ENDODONTIC, IRRIGATING Back to Search Results
Catalog Number V041441000000
Device Problems Break (1069); Inaccurate Synchronization (1609)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude injury or illness that would necessitate medical or surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.This event, therefore, is reportable per 21cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.
 
Event Description
In this event a customer reported that an eddy irrigation tip broke; no injury resulted.
 
Manufacturer Narrative
1st tips with cavity 2; tips is missing 13.3mm measured from 28mm point.2nd tip cavity 4, broken in 2 pieces - nothing missing.Measuring gauge mitutoyo model # cd 15cpvalid to 5/2020.No smooth breakage, melted.Breakage due to thermal influence - misuse.Test of unused product passed the safety and mechanical test.No fault found on unused instruments.Nothing unusual to report was found during dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDDY IRRIGATION TIP
Type of Device
SYRINGE, PERIODONTIC, ENDODONTIC, IRRIGATING
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
MDR Report Key8596905
MDR Text Key144846503
Report Number9611053-2019-00129
Device Sequence Number1
Product Code EIC
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041441000000
Device Lot Number244108
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2019
Date Manufacturer Received06/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-