Catalog Number V041441000000 |
Device Problems
Break (1069); Inaccurate Synchronization (1609)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
As a result of this malfunction, the potential for surgical intervention exists to preclude injury or illness that would necessitate medical or surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function.This event, therefore, is reportable per 21cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.
|
|
Event Description
|
In this event a customer reported that an eddy irrigation tip broke; no injury resulted.
|
|
Manufacturer Narrative
|
1st tips with cavity 2; tips is missing 13.3mm measured from 28mm point.2nd tip cavity 4, broken in 2 pieces - nothing missing.Measuring gauge mitutoyo model # cd 15cpvalid to 5/2020.No smooth breakage, melted.Breakage due to thermal influence - misuse.Test of unused product passed the safety and mechanical test.No fault found on unused instruments.Nothing unusual to report was found during dhr review.
|
|