Model Number V60
Device Problem Material Discolored (1170)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 13may2019.A follow-up report will be submitted once the investigation has been complete.
 
Event Description
The customer reported that the color of the background turned to orange.The device was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or user.
 
Manufacturer Narrative
Date rec¿d by mfr : 20jun2019, date of report : 26jun2019.The manufacturer¿s international service technician could not duplicate the reported problem.The service technician replaced the lcd preventively.Since the lcd changed to the 2nd gen model, the user interface (ui) pcba, the 2nd gen lcd cable, the ui pcba-lcd cable and the 2nd gen lcd tray were replaced.The unit was checked overall, run in tested, cleaned and functionally tested and no abnormality was confirmed.There is a relationship of the device to the reported problem.The root cause cannot be determined until the device is returned and investigated.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8605081
MDR Text Key144967091
Report Number2031642-2019-02949
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-