Model Number SN6AT3
Device Problem Positioning Problem (3009)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 02/21/2019
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
 
Event Description
A nurse reported that following an intraocular lens (iol) implant procedure, the astigmatism was not reduced.A secondary procedure was performed to reposition the iol from 90 degrees to 180 degrees.The doctor reported that there was an axis problem when the lens was manufactured.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Type of Device
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key8608141
MDR Text Key145023613
Report Number1119421-2019-00678
Device Sequence Number1
Product Code MJP
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberSN6AT3
Device Catalogue NumberSN6AT3.220
Device Lot Number12620737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-