The automated manufacturing execution system (mes) has controls in the manufacturing process to ensure the product met specifications upon release.The device history record review confirms that the device met all material, assembly and performance specifications.The stent was returned with the stryker microcatheter used during the procedure.Visual and microscopic inspection of the device found that the stent delivery wire (sdw) was broken at the distal bumper and the stent was found to be deployed inside the microcatheter hub along with the distal section of the sdw.The stent was noted to be deformed as well.Functional test could not be performed due to the damaged condition of the returned device.The stent was returned prematurely deployed within the stryker xt-17 microcatheter hub.It is likely that procedural factors contributed to the reported and observed damages limiting the performance of the stent during the clinical procedure.Therefore, an assignable cause of component failure has been assigned to this investigation.
|