Manufacturer report number: 2938836-2019-03733, manufacturer report number: 2938836-2019-03734.It was reported patient presented in the emergency room when noise on both the right atrial and right ventricular leads were noted.It was observed patient had received more than forty inappropriate therapies due to lead noise.The device was programmed doo and tachycardia therapies were disabled.The following day, the device was interrogated again, and the same noise was observed.Low pacing impedance was also noted on the right atrial, right ventricular and left ventricular leads.A chest x-ray was taken, and physician suspected lead abrasion was likely the cause of the noise and impedance drop.On (b)(6) 2019 just prior to procedure, the device was interrogated and low out of range high voltage (hv) lead impedance was observed.The right ventricular lead was explanted and replaced.The atrial lead was left in place and plugged in as the patient was chronic atrial fibrillation and device was programmed to vvi.During the procedure, pacemaker system analyzer testing on the left ventricular lead showed normal impedance and pacing characteristics and the lead remained implanted.There were no complications and patient was stable throughout.
|
Related manufacturer reference number: 2938836-2019-13294 new information received notes, on arrival to the emergency, once in the room patient had potential episode of vt/vf, syncope and was reportedly pulselessness.He received chest compression and external shock to obtain return of pulses.Noise was also noted on the left ventricular lead.Prior to the procedure, the physician was concerned that the patient¿s device was not working correctly given the amount of interference and noise from the leads.The device was also explanted and replaced.
|