HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC Back to Search Results
Model Number P3700D000015
Device Problem Unintended System Motion (1430)
Patient Problem No Patient Involvement (2645)
Event Date 05/06/2019
Event Type  malfunction  
Manufacturer Narrative
Hill-rom technical support suggested replacing the right caregiver side rail cable.Per the hill-rom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the switches in the side rails to ensure they are functioning correctly.Also check for intermittent operation.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account will replace the right caregiver side rail cable to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head section self ran up.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
TABLE, OBSTETRIC
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key8615822
MDR Text Key145538094
Report Number1824206-2019-00208
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberP3700D000015
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/06/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/18/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-