BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS Back to Search Results
Catalog Number 762165
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Information (3190)
Event Date 05/02/2019
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that 35 paxgene® blood rna tubes experienced discoloration/abnormal additive form.The following information was provided by the initial reporter: customer observed turbidity in the tubes while using and reported to us.Turbidity in paxgene rna tube.
 
Event Description
It was reported that 35 paxgene® blood rna tubes experienced discoloration/abnormal additive form.The following information was provided by the initial reporter: customer observed turbidity in the tubes while using and reported to us.Turbidity in paxgene rna tube.
 
Manufacturer Narrative
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for turbidity with the incident lot was observed.Additionally, retention samples were selected from bd inventory for evaluation/testing and upon completion, the issue relating to turbidity was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for turbidity with the incident lot was observed.Evaluation/testing of the retain samples was also conducted and turbidity was not observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAXGENE® BLOOD RNA TUBE
Type of Device
RNA PREANALYTICAL SYSTEMS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
MDR Report Key8617524
MDR Text Key145521785
Report Number9617032-2019-00602
Device Sequence Number1
Product Code NTW
Combination Product (y/n)N
PMA/PMN Number
DEN050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 06/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2019
Device Catalogue Number762165
Device Lot Number8081579
Date Manufacturer Received05/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-