Model Number AB-5100R |
Device Problem
Premature Discharge of Battery (1057)
|
Patient Problems
Headache (1880); Failure of Implant (1924)
|
Event Date 03/13/2019 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly elected explant surgery due to poor battery life.The recipient presented with headaches.The recipient's device was explanted.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the fantail was damaged near the epoxy header, and the electrode was sliced along the lead prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection confirmed broken electrode wires at the epoxy header.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.Voltage values at some nodes were elevated.Internal inspection did not reveal any anomalies on internal components.This older device configuration is not currently manufactured.This is the final report.
|
|
Manufacturer Narrative
|
The recipient was reimplanted with another cochlear device.Additional information regarding recipient status will not be provided.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|