• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN LINER; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN LINER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Scratched Material (3020)
Patient Problem Osteolysis (2377)
Event Date 04/10/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: unknown stem, unknown hg cup, unknown head.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2019-02105, 0001822565-2019-02147.Radiograph was provided and reviewed by a health care professional.Review of the available records identified eccentric position of the femoral head within acetabular cup reflecting advanced poly wear.Multiple foci of abnormal radiolucency consistent with osteolysis of the femoral component and at multiple areas along acetabular component were present.No further evaluation was possible with the images provided.Reported event was confirmed by review of x-rays provided.Device history record review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Device remains implanted in the patient.
 
Event Description
It was reported that the patient was revised due to unknown reasons approximately 25 years post implantation.Attempts have been made and additional information on the reported event is unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8627177
MDR Text Key145613750
Report Number0001822565-2019-02146
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKNOWN LINER
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight59
-
-