Model Number 1458Q/86
Device Problem Misconnection (1399)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During device replacement for normal battery depletion, it was reported that the left ventricular (lv) lead could not be removed from the header of the device.The lv lead was capped and replaced to resolve the event and the device was explanted and replaced.The patient was stable and there were no adverse consequences.Related manufacturer reference number: 2938836-2019-03980.
 
Manufacturer Narrative
The reported event was lead stuck in header.As received, a partial lead was returned in two pieces.Electrical testing did not find any indication of conductor fractures or internal shorts.Dimensional analysis was performed on the ring electrodes which measured within specification.Visual inspection of the lead found no anomalies with the exception of damages consistent with procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8629837
MDR Text Key145741635
Report Number2017865-2019-08473
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Model Number1458Q/86
Device Lot Number3914214
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CD3267-40, (B)(4)
-
-