SURE-T® PARADIGM®; CONTACT DETACH G29 80/10 PCC Back to Search Results
Lot Number 5221262
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Tissue Damage (2104)
Event Type  malfunction  
Event Description
Unomedical reference number (b)(4).Event occurred in the united states.The patient's father reported that the infusion set was inserted, and manual injection was given to the patient.After two days of infusion set use, the patient had lunch and his blood glucose level started to go up.His blood glucose level was 550 mg/dl when the father took the old infusion set off to replace it.It was found that the needle was not attached to the set and the entire needle broke off into the patient's body.The patient was taken to the emergency room and was subsequently, admitted to the hospital to get the needle removed with blood glucose level at around 120 mg/dl.The patient received medical treatment in the form of a surgery to remove the needle that had fractured inside the patient's body.The infusion set insertion location was in upper butt area.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURE-T® PARADIGM®
Type of Device
CONTACT DETACH G29 80/10 PCC
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key8630576
MDR Text Key145742928
Report Number3003442380-2019-01485
Device Sequence Number1
Product Code FPA
UDI-Device Identifier05705244006034
UDI-Public05705244006034
Combination Product (y/n)N
PMA/PMN Number
K160648
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date01/01/2021
Device Lot Number5221262
Date Manufacturer Received04/17/2019
Type of Device Usage N
Patient Sequence Number1
-
-