• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-60-00
Device Problem Calibration Problem (2890)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Model number and serial number are unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova (b)(4)manufactures the s5 roller pump.The incident occurred in (b)(6).The livanova field service representative in charge replaced the potenziometer and was thus able to remove the issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the potentiometer of a s5 roller pump was out of tolerance.The issue was identified during the device maintenance.There was no patient involvement.
 
Event Description
See initial.
 
Manufacturer Narrative
H.10: the model number and the serial number have been added to the d4 section.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8636012
MDR Text Key145950088
Report Number9611109-2019-00377
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-60-00
Was the Report Sent to FDA? No
Date Manufacturer Received06/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-