Catalog Number 306594
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Information (3190)
Event Date 05/20/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: one photo was provided.It shows a syringe with the barrel flange damaged.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: confirmed: bd was able to duplicate or confirm the customer's indicated failure mode with the photo provided.This is the 1st complaint for lot # 8045562 for this type of defect or symptom.There was no documentation for this type of defect during the entire production run of this batch.Root cause description: the damaged barrel flange is usually is from the plunger rod labeler process.Rationale: capa# (b)(4) has been initiated to address this issue.
 
Event Description
It was reported that during use, 2 bd posiflush¿ syringes were found with damaged barrel flanges that resulted in sharp edges.Capa# (b)(4) was opened in response to this event.The following information was provided by the initial reporter, translated from chinese to english: "it was found the end of barrel damaged.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key8653876
MDR Text Key146985389
Report Number1911916-2019-00534
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/28/2021
Device Catalogue Number306594
Device Lot Number8045562
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-