• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, 30CM; SCS EXTENSION Back to Search Results
Model Number 3383
Device Problem Disconnection (1171)
Patient Problems Fall (1848); Inadequate Pain Relief (2388)
Event Date 05/14/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Reference mfr report: 3006705815-2019-02022.It was reported that patient experienced stimulation lost from the cervical system after a fall.Reportedly, x-rays confirmed that the extension had disconnected from the ipg header.As a result, patient underwent surgical intervention on (b)(6) 2019 wherein the extension and ipg were explanted and replaced.Effective therapy was restored post operatively.
 
Event Description
Reference mfr report: 3006705815-2019-02022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, 30CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8659177
MDR Text Key146678195
Report Number1627487-2019-06338
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05414734402323
UDI-Public05414734402323
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960728
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/03/2019
Device Model Number3383
Device Lot Number5967949
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/02/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3772, SCS IPG
Patient Outcome(s) Other;
Patient Weight75
-
-