Model Number 53401
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 08/14/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis noted the external pulse generator (epg) failed incoming testing, the main printed circuit board (pcb) was out of electrical specification.All found defective parts were replaced and the firmware was updated.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The external pulse generator was returned to the company service department for regular preventative maintenance and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: further analysis of the printed circuit board (pcb) determined that the pcb was in specification.There were no anomalies and no defects noted.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPORARY EXTERNAL PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8659291
MDR Text Key146682574
Report Number3004593495-2019-00508
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169723160
UDI-Public00643169723160
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number53401
Device Catalogue Number53401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2019
Date Manufacturer Received07/01/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-