Model Number 53401 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis noted the external pulse generator (epg) failed incoming testing, the main printed circuit board (pcb) was out of electrical specification.All found defective parts were replaced and the firmware was updated.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The external pulse generator was returned to the company service department for regular preventative maintenance and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
|
|
Manufacturer Narrative
|
Product analysis: further analysis of the printed circuit board (pcb) determined that the pcb was in specification.There were no anomalies and no defects noted.If information is provided in the future, a supplemental report will be issued.
|
|