WRIGHTS LANE SYNTHES USA PRODUCTS LLC 8MM TI CURVED RADIAL STEM 44MM-STERILE; PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER Back to Search Results
Catalog Number 04.402.028S
Device Problem Unintended Movement (3026)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Weakness (2145)
Event Type  Injury  
Manufacturer Narrative
This report is for an unknown radial stem /unknown lot.Part and lot numbers are unknown; udi number is unknown.Initial reporter is synthes legal representative without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2018, the patient underwent revision surgery due to the loosening of the radial head prosthesis which the patient experience increasing pain, weakness, loss of motion and decreased the function of the elbow.Originally, the patient underwent open reduction internal fixation for radial head replacement on (b)(6) 2016, due to a 4-part comminuted right radial head fracture.The patient was treated with depuy synthes radial head system.The surgical delay was unknown.The procedure outcome was successfully completed, and the patient status was stable.This report is for one (1) radial head prosthesis: radial stem.This is report 1 of 2 for (b)(4).
 
Event Description
Updated event description: it was reported that on (b)(6) 2018, the patient underwent revision surgery due to the loosening of the radial head prosthesis which the patient experience increasing pain, weakness, loss of motion and decreased the function of the elbow.The screw was loosened on the synthes radial head implant and the head was removed.The stem was loose and was removed as well.Imaging demonstrated widening and scalloping of the proximal radius with extreme thinning of the lateral cortex.Procedure was done to maintain radial anatomy, specifically pro/supination strength and avoiding malunion, painful nonunion reportedly, the patient underwent open reduction internal fixation for radial head replacement on (b)(6) 2016, due to a 4-part comminuted right radial head fracture.The patient was implanted with a synthes radial head and stem.Postoperatively, the patient underwent a left knee arthroscopy and partial medial menisectomy/excisional biopsy ganglion cyst on (b)(6) 2017 due to left knee and swelling.Exam and imaging shows tear of the medial meniscus with a large ganglion cyst near pes anserine bursa that was communicating with the medial compartment.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Patient's height is reported as 6'0".Customer quality investigation: the complaint device was not received for investigation.The pictures were reviewed and based on the pictures alone us cq can confirm the complaint.The implant appears to have loosened in the implant site as seen in image.This is contrasted with another image where the implant is in the correct position.There are known issues with the radial head implants that are captured in corrective and preventative action (capa).There is also a recall (recall# 555531) that was initiated as part of field action investigation where ¿device loosening¿ was identified as a high harm and contributed to the recall of the product.Based on the images, capa, hhe and field action information us cq can confirm the complaint for loosening.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A definitive assignable root cause could not be determined based on the provided information.During the investigation no unidentified product design/manufacturing issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.However, based on previous complaints, the radial head prosthesis system (rhp system) was recalled under recall# 555531.Any related investigations and assessment of the risks associated with this system will be covered under recall# 555531 and hhe-2016-180 investigations.Since the radial head prosthesis system has been recalled and the complaint condition was investigated through those efforts, no further investigation is required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.A corrective and/or preventative action is proposed or already launched.See related actions a device history record (dhr) review was conducted: manufacturing location: supplier - avalign nemcomed / inspected, packaged and released by: monument, release to warehouse date: 02-jul-2012, expiration date: 01-may-2017, part number: 04.402.028s, 8mm ti curved radial stem 44mm ¿ sterile, lot number: 6970383 (sterile) ¿ pulled from lot number: 6873703.Lot quantity: (b)(4).Work order travelers met all inspection acceptance criteria.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Component part(s) reviewed: part number: 21069, tialnbri18.00, lot number: 6737206, lot quantity: (b)(4).This lot met all dimensional, visual, sterility and packaging criteria with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8MM TI CURVED RADIAL STEM 44MM-STERILE
Type of Device
PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8660212
MDR Text Key146719347
Report Number2939274-2019-58401
Device Sequence Number1
Product Code KWI
Combination Product (y/n)N
PMA/PMN Number
K112030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2017
Device Catalogue Number04.402.028S
Device Lot Number6970383
Was Device Available for Evaluation? No
Date Manufacturer Received10/25/2019
Removal/Correction NumberZ-1124-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient Weight85
-
-