ABBOTT CRM QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1258T/86
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined. .
 
Event Description
During follow-up, oversensing due to noise was noted on the left ventricular lead.The device was reprogrammed to resolve the event.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT CRM
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT CRM
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8660763
MDR Text Key146908245
Report Number2017865-2019-08981
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Model Number1258T/86
Device Lot Number3755866
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/19/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-