Model Number 48VMP160
Device Problems Leak/Splash (1354); Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/14/2019
Event Type  malfunction  
Manufacturer Narrative
One vamp adult kit was returned for examination.The reported event of ¿saline leakage from stopcock ¿was confirmed.As received the tubing had detached from the bond joint with a vamp adult reservoir stopcock.The tubing was bent near the point of detachment.Indications of bonding solvent were evident on the tubing bond surface area.The tubing outer diameter was measured near the point of detachment and was found to be within specification.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.It is common clinical practice to inspect all products before usage.These products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise during use.In addition, they are used in critical care units or ors where patients are closely monitored.If the pressure tubing becomes detached during use, it will affect the pressure waveform, which will immediately alert the clinician to begin the troubleshooting process.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.In this event, no patient compromise was reported as the issue was noted before use.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that before use of this dpt (disposable pressure transducer), saline leaked from the stopcock.Replacing the device resolved the issue.There was no allegation of patient injury.The device was available for evaluation.
 
Manufacturer Narrative
The investigation of this vamp adult kit concluded that a potential root cause could be related to an incorrect solvent bonding execution during the assembly process.Personnel acknowledgement was conducted at the manufacturing site to prevent recurrence of this type of complaint.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAMP VENOUS ARTERIAL BLOOD MANAGEMENT PROTECTION SYSTEM WITH ARM MOUNT RESERVOIR
Type of Device
KIT, SAMPLING, ARTERIAL BLOOD
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key8668563
MDR Text Key147321286
Report Number2015691-2019-01992
Device Sequence Number1
Product Code CBT
Combination Product (y/n)N
PMA/PMN Number
CLASS I EXEM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/29/2020
Device Model Number48VMP160
Device Lot Number61386314
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2019
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Weight73
-
-