• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. OVERSTITCH ENDOSCOPIC SUTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. OVERSTITCH ENDOSCOPIC SUTURING SYSTEM Back to Search Results
Model Number ESS-G02-160
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problems Internal Organ Perforation (1987); Pain (1994)
Event Date 05/01/2019
Event Type  Injury  
Manufacturer Narrative
The physician indicated the device will not be returned to apollo for analysis.A review of the device labeling notes the following: warning: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.Users should be familiar with surgical procedures and techniques involving absorbable sutures before employing synthetic absorbable sutures for wound closure, as the risk of wound dehiscence may vary with the site of application and the suture material used.Ensure that there is sufficient space for the needle to open.Warning: do not introduce the device with the needle body in its open position.Adverse event: possible complications that may result from using the endoscopic suturing system include, but may not be limited to: abdominal pain and / or bloating.Pharyngeal, colonic and/or esophageal perforation.Esophageal, colonic and/or pharyngeal laceration.Intra-abdominal (hollow or solid) visceral injury.
 
Event Description
Reported as: a patient who underwent an endoscopic sleeve gastroplasty utilizing the overstitch endoscopic suturing system was noted to have a complication.A patient had a procedure and was doing well however they began to have pain while standing up.The patient went to the er and the trauma surgeon did laparotomy and found a gastric perforation.The physician took down the esg and repaired the defect.Notes from procedure indicate a "full thickness tear from a full thickness suture that pulled through." physician believes it was the reinforcement stitch and it was cinched too tightly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVERSTITCH ENDOSCOPIC SUTURING SYSTEM
Type of Device
SUTURING SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX 78746
Manufacturer (Section G)
VIANT MEDICAL
5079 33rd street se
grand rapids MI 49512
Manufacturer Contact
laura leboeuf
1120 s. capital of texas hwy
bldg 1, ste. 300
austin, TX 78746
MDR Report Key8670412
MDR Text Key147051387
Report Number3006722112-2019-00124
Device Sequence Number1
Product Code OCW
UDI-Device Identifier10811955020664
UDI-Public10811955020664
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberESS-G02-160
Device Catalogue NumberESS-G02-160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/05/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age36 YR
-
-