Catalog Number 826653 |
Device Problem
Device Sensing Problem (2917)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that the microsensor could not be detected by the icp and then the physician changed with another icp which still could not detect the microsensor.Therefore the problem was with the microsensor.The physician stopped monitoring.The event occurred pre operatively; there were no delays and no adverse consequences.The device will be returned.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Udi (b)(4).The catheter was evaluated and the following observations were noted.Received catheter in drainage tube, the internal wires were broken(x-ray).The adhesive was dirty or the sensor area.No testing was possible.Device history records (dhrs) were reviewed and no anomalies were found.The complaint failure has been confirmed.The root cause was due to mishandling of the catheter.No further investigation or corrective action is anticipated.
|
|
Search Alerts/Recalls
|