Evaluation results: one level 1 hotline low flow system (hl-90) was returned for investigation in used condition.The device was received with scratches to the cover, and some discoloration was observed on the tank.The device was manufactured in 2008.The investigator powered on the unit and it alarmed check disposable.The investigator then moved the switch into a normally closed position and powered on the device again.The alarm activated when it should not have.The investigator removed the cover for further inspection.The investigator subsequently measured the open circuit in both normally open and normally closed states.This revealed that the float switch was faulty.The investigator then installed a known good test switch and performed a functional check.The disposable alarm then functioned as designed.The device was scrapped due to age however.
|