MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA#HLS SET ADVANCED 7.0
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).The product was requested for return to the manufacturer for laboratory investigation but was not received yet.The investigation is still pending.A supplemental medwatch will be submitted after new information has been received.
 
Event Description
It was reported that hls set was stopped reading pressures during a support run.The device was not function properly.The set was replaced with a new.No harm to the patient was reported.After follow up with the ccp, there appears to be green residue within the cable that attaches the hls set to the cardiohelp-i console.The issue with the cable will be handled in a separate complaint ((b)(4)).(b)(4).
 
Manufacturer Narrative
Maquet cardiopulmonary gmbh requested the product for investigation on (b)(6)2019.This complaint will be closed since the requested part has not been returned for investigation despite several requests.The complaint will be re-opened if requested part or any new relevant information is received.Thus the failure could not be confirmed the ocurrance rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint number:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key8678057
MDR Text Key147319865
Report Number8010762-2019-00162
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBEQ-HLS 7050 USA#HLS SET ADVANCED 7.0
Device Catalogue Number701052794
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-