Model Number 10-80-00
Device Problem Failure to Power Up (1476)
Patient Problem No Patient Involvement (2645)
Event Date 03/29/2019
Event Type  malfunction  
Manufacturer Narrative
This event was initially considered to be non-reportable.However, after additional evaluation, livanova (b)(4) has decided to reclassify this event as reportable.This report is being filed as part of a retrospective review conducted in response to this new decision.There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a on-/off switch.The technician replaced the switch to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.As the switch was not made available no further investigation could be performed and no specific root cause could be determined.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 roller pump would not stay powered up during setup.There was no patient involvement.
 
Manufacturer Narrative
The affected part was returned to livanova deutschland for a detailed investigation.The investigator could reproduce the reported issue.The problem could be addressed to an electronic component problem.Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8678270
MDR Text Key147327223
Report Number9611109-2019-00440
Device Sequence Number1
Product Code DWB
UDI-Device Identifier04033817900382
UDI-Public(01)04033817900382(11)161014
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was the Report Sent to FDA? No
Date Manufacturer Received07/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-