The device history record review confirms that the device met all material, assembly and performance specifications.Visual and microscopic inspection of the device found that the proximal end of the stabilizer was broken, distal portion of the catheter was severely flattened and the catheter was kinked from the distal end.The stent was returned in its deployed state (physician deployed the stent post procedure) with some deformation noted.No other anomalies were noted.During functional test, after flushing, the stent stabilizer was unable to be advanced or retracted from the catheter due to the damage observed.It is likely that procedural factors contributed to the reported and observed damages limiting the performance of the stent during the clinical procedure.Therefore, an assignable cause of operational context has been assigned to this investigation.
|