Model Number NM-401L-0425
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 05/22/2019
Event Type  Injury  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, because the subject device was discarded by the user.The exact cause has been under investigation.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event Description
During a steroid injection for subglottic stenosis, the subject device was used.The user could not push a steroid through the needle tube regardless of amount of a pressure.The user had to inject a steroid by trans-tracheal.The subject device was discarded.No further information was provided.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, because the subject device was discarded by the user.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the event occurred due to compressive buckling of the needle tube.The compressive buckling might occur since the friction resistance between the outer tube and the needle tube was high when the needle tube was pushed out.The friction resistance might be high due to the following factors.During inspection, the needle extended or retracted while the insertion portion was coiled.The insertion portion was bent tightly during use.The slider was pushed abruptly.The tube was buckling.The end of the endoscope was angulated.The above device handling has warned in the instruction manual as follows.Straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Type of Device
INJECTOR AND SHEATHSET
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8681994
MDR Text Key147439868
Report Number8010047-2019-02156
Device Sequence Number1
Product Code FBK
UDI-Device Identifier04953170260070
UDI-Public04953170260070
Combination Product (y/n)N
PMA/PMN Number
K902736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 08/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNM-401L-0425
Device Lot Number84K
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-