In (b)(6) 2016 a patient received a carbomedics aortic mechanical valve, size 19.The manufacturer was notified that during the patients follow-up (approx.(b)(6) 2018) lv systolic function is within normal limits, but the echo doppler gradient across the aortic valve has increased, now pipg 52, mean 25 mm hg.On (b)(6) 2019 the manufacturer was notified echo shows a minimal increase in doppler gradient (pipg 55, mean 32, compared to six months ago pipg 52, mean 25), with normal lv size and function.They also identified the posterior leaflet was not moving normally, suspected by echo, and confirmed by fluoroscopy.His inr is 1.9.The child is not meeting criteria for surgical reop on his aortic valve per site decision.
|