• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND SCREW LOCK D4.5X36MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND SCREW LOCK D4.5X36MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Catalog Number 130790036
Device Problem Loss of Osseointegration (2408)
Patient Problems Inadequate Osseointegration (2646); No Code Available (3191)
Event Date 05/28/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient was revised to address fall and loosening of the metaglene at cement to implant interface.It was also reported that patient likely broke 2 screws from the fall.Surgeon decided to revise patient shoulder and removed the glenospheres, the metaglene base plate, 4 screws, and the humeral cup.The glenoid was filled with bone graft and a delta cta head was implanted onto stem.Doi: (b)(6) 2012; dor: (b)(6) 2019; left shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X36MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
Manufacturer (Section G)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
b.p. 256
leeds 69801
UK   69801
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, IN 19380-0988
6103142063
MDR Report Key8703684
MDR Text Key148159617
Report Number1818910-2019-96104
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027850
UDI-Public10603295027850
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130790036
Device Lot Number5200168
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age54 YR
-
-