• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASPEN SURGICAL PRODUCTS, CALEDONIA RICHARD ALLAN MODERN REG TIP PEN W/ RULER & 6-UP LABELS; MARKING PEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASPEN SURGICAL PRODUCTS, CALEDONIA RICHARD ALLAN MODERN REG TIP PEN W/ RULER & 6-UP LABELS; MARKING PEN Back to Search Results
Model Number 2750
Device Problem Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/28/2019
Event Type  malfunction  
Manufacturer Narrative
Aspen surgical received a report from the distributor that product was found with seal issues.The actual device was returned for evaluation.The manufacturing lot number was provided for review as well.The return of the sample confirmed the issue identified from the distributor.Analysis of the finished good lot number was reviewed.No non-conformance's were noted during the manufacturing process.Upon review of the samples, the seals on the film show incomplete transfer of adhesive.This incomplete seal can be caused by unexpected process interruptions identified in operating instructions.Additionally, according to the manufacturing process, these parts are manually loaded into recessed pockets on a packaging machine per specified work instructions.If incorrectly loaded or not orientated properly, this will cause an incomplete seal.Therefore a likely root cause for the seal issue may be attributed to an operator error.The pouch label, identifies this failure mode with the symbol "do not use if package is damaged".This indicates that the device should not be used if the products sterile barrier system or its packaging is compromised.Additionally, production supervisors and operation were notified of this issue during the daily update meetings.Based on this information, no further action is required.
 
Event Description
Aspen surgical received a report from the distributor indicating that richard-allan regular tip pen was found with defective seals.The item was not in use.No injury/death was reported.This report was filed in our complaint handling system as number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RICHARD ALLAN MODERN REG TIP PEN W/ RULER & 6-UP LABELS
Type of Device
MARKING PEN
Manufacturer (Section D)
ASPEN SURGICAL PRODUCTS, CALEDONIA
6945 southbelt dr. s.e.
caledonia MI 49316
Manufacturer (Section G)
SAME AS ABOVE
Manufacturer Contact
jordan hackert
6945 southbelt dr. s.e.
caledonia, MI 49316
6165367508
MDR Report Key8707378
MDR Text Key151427187
Report Number1836161-2019-00050
Device Sequence Number1
Product Code FZZ
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2750
Device Lot Number179092
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-