Catalog Number 8065977763 |
Device Problems
Break (1069); Fracture (1260)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There has been one other complaint reported in the lot number.(b)(4).
|
|
Event Description
|
An ophthalmic technician reported that during an intraocular lens (iol) implant procedure, the cartridge split during implantation.A stress fracture was also noted on the cartridge nozzle.The procedure was completed with a new cartridge.There was patient contact, but no patient harm.
|
|
Manufacturer Narrative
|
A used cartridge was returned in the opened pouch.Viscoelastic is observed in the cartridge.The nozzle is cracked on the right side.This damage turns into a split as it enters the thinner tip area.The cartridge has evidence it was placed into a handpiece.The cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and the documentation indicated the product met release criteria.The provided associated products are qualified for use with the cartridge.The root cause for the tip damage could not be determined.The returned used cartridge nozzle is cracked on the right side.This damage turns into a split as it enters the thinner tip area.The damage to the nozzle started in the thick wall cone area of the nozzle.Unusually high internal forces would be needed to create damage in this area.The two distinct areas of damage would indicate a progressive change that occurred as the lens was advanced.This type of damage typically occurs if the lens is not positioned correctly for advancement; if there is a lack of viscoelastic between the lens and the cartridge lumen; if the lens is advanced too rapidly or if the handpiece plunger is not positioned correctly at the trailing optic edge.If the handpiece plunger is not positioned at the trailing optic edge, it can allow the lens to fold around the plunger tip making it too large to correctly advance through the narrow tip of the cartridge, which could cause damage to the tip or the lens.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|