Model Number US751
Device Problems Contamination (1120); Material Discolored (1170)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2019
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with the round filters.There are some ink spots noted on the back of the filters before sterilization.It appeared to be the same color as the indicator dot.After the filters were sterilized there are more spots that appeared.There was no patient harm or surgery reported.
 
Manufacturer Narrative
Additional information - manufacturing site evaluation: the device was not returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation of the unit was performed and no parts were returned for failure analysis.Therefore, the investigation was not able to confirm a device issue that could be associated with the reported event.The device history records (dhr) were reviewed for all available lot numbers; the devices were found to be within specification at the time of production.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROUND FILTERS W/INDICATOR
Type of Device
STERILIZATION WRAP CONTAINERS, TRAYS, CASSETTES & OTHER ACCESSORIES
Manufacturer (Section D)
AESCULAP, INC.
3773 corporate parkway
center valley PA 18034
MDR Report Key8713688
MDR Text Key148549261
Report Number2916714-2019-00027
Device Sequence Number1
Product Code KCT
Combination Product (y/n)N
PMA/PMN Number
K944864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUS751
Device Catalogue NumberUS751
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date05/22/2019
Date Manufacturer Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-