• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number ASK-05500-KM
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/27/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the user found it difficult to insert a catheter into the epidural needle because of resistance.The kit was replaced with a new one.
 
Event Description
It was reported that the user found it difficult to insert a catheter into the epidural needle because of resistance.The kit was replaced with a new one.
 
Manufacturer Narrative
Qn#(b)(4).A device history record review was performed on the epidural catheter and snaplock assembly with no relevant findings.The customer reported the medical agent would not flow through the catheter.The customer returned one empty kit with a snaplock assembly and an epidural catheter.The returned components were visually examined with and without magnification.Visual examination of the returned snaplock assembly revealed the snaplock appears typical with no observed defects or anomalies.Visual examination of the returned catheter revealed that the catheter appears typical used as biological material can be seen in the inner coils.No other defects or anomalies were observed.A functional flow test was performed on the returned sample per amrq-000017 section 7.8; rev.7.The returned epidural catheter was inserted from the proximal end into the returned snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester (c05176) and the pressure was increased to 10 psi to establish flow.Water could be seen immediately exiting the distal end of the catheter.The flow rate was measured at 8.4ml/min (c05180), which is within the specification of 1ml/min.No blockages were found.The reported complaint of the medical agent not flowing through the catheter could not be confirmed based on the sample received.A device history record review was performed on the epidural catheter and snaplock assembly with no evidence to suggest a manufacturing related issue.The returned components passed a functional flow test and met flow rate specifications.There were no functional issues found with the returned sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8713800
MDR Text Key148538107
Report Number3006425876-2019-00419
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2020
Device Catalogue NumberASK-05500-KM
Device Lot Number71F18L0458
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2019
Date Manufacturer Received07/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-