Model Number 16414 |
Device Problem
Use of Device Problem (1670)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported that there was an issue with their cardioplegia monitor.No other details regarding the nature of this event were provided.
|
|
Event Description
|
Additional information received that the monitor would not come on and the issue occurred during setup of the device for a cardiopulmonary bypass (cpb) procedure.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint could not be confirmed.The field service representative (fsr) was not able to duplicate the reported complaint.The cardioplegia (cpg) monitor powered on and functioned properly.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|