It was reported that during a tlif spine procedure, the cautery device tip was identified to be flaking.When this flaking was identified the cautery device was removed from the field and the tip was replaced.The procedure continued without further reported incident.Reportedly, the flakes were "so small you can't see it." no irrigation or removal of the flakes occurred.There was no reported impact or adverse effect to the patient.A sample of only the cautery device tip was returned to the manufacturer for evaluation.Visual inspection of the sample confirmed the reported flaking of the cautery device tip.No information was submitted in regard to the settings and use of the cautery of device or the settings and type of generator used at the time of the incident.A root cause for the reported incident could not be determined.Due to the reported incident, and in an abundance of caution, this medwatch is being filed.If additional relevant information becomes available a supplemental medwatch will be filed.
|