• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL Back to Search Results
Model Number RD900
Device Problem Device Handling Problem (3265)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint rd900 neopuff infant resuscitator was not returned to the fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the initial information reported by the customer.The customer reported that the peak inspiratory pressure (pip) valve's adjustment knob of the complaint rd900 neopuff infant resuscitator is broken.We are unable to determine what has caused the reported event without the complaint device.The neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to an infant until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to damage, for instance when accidentally dropped or subjected to considerable external force.The physical damage reported was most likely due to impact to the subject neopuff unit.It must also be noted that the subject neopuff is more than 12 years old.The neopuff technical manual states the following: "dropping the neopuff infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient." each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.
 
Event Description
A hospital in (b)(6) reported that a peak inspiratory pressure (pip) valve's adjustment knob of a rd900 neopuff infant resuscitator is broken.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOPUFF INFANT RESUSCITATOR
Type of Device
BTL
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9194534000
MDR Report Key8725611
MDR Text Key149132689
Report Number9611451-2019-00607
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K892885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRD900
Device Catalogue NumberRD900
Device Lot Number061109
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/25/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/09/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-