The device was investigated at emtec with the rma# 2019-10187 on the 2019-09-23: problem description: error head.Work performed: ild (optocoupler)and boards mc1 and mc2 replaced.All tests passed.Root cause: hot plug.The device was also investigated withe rma# 38785 and the maquet service order# 10477727 on the 2019-09-26: could be confirmed: ild (optocoupler)and boards mc1 and mc2 replaced.System test performed according to service protocol.In the course of the investigation of (b)(4) all complaints were analyzed individually to look for patterns and causes.After evaluation of the complaints, the following defects are the most common: hot plug, sig error followed by head error, error message due to shaking / error message due to sensitivity, connection problems and hardware-error.The increase in complaints with the error "head error" is due to a user / application error.The actions have been addressed in the past to prevent application errors.Furthermore, the instructions for use of the rotaflow system see rotaflow system user manual, (b)(4), contain detailed descriptions to prevent an "error head".In addition, the instruction manual describes how the user has to react in case of an error message so that the application can be continued if possible.Since there are several causes of errors that result in an error head (error head) message, no final root cause could be determined.It is noticeable that the evaluated error is largely due to a user error.Warning in the ifu regarding "hotplug" and label with the "hotplug" warning on the rfc have already been implemented in the past.
|