• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number ROTAFLOW DRIVE UNIT , BLUE
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device not returned not eval.
 
Event Description
The rotaflow showed the error message "head error".No patient involvement.Complaint #: (b)(4).
 
Event Description
Complaint #: (b)(4).
 
Manufacturer Narrative
The device was investigated at emtec with the rma# 2019-10187 on the 2019-09-23: problem description: error head.Work performed: ild (optocoupler)and boards mc1 and mc2 replaced.All tests passed.Root cause: hot plug.The device was also investigated withe rma# 38785 and the maquet service order# 10477727 on the 2019-09-26: could be confirmed: ild (optocoupler)and boards mc1 and mc2 replaced.System test performed according to service protocol.In the course of the investigation of (b)(4) all complaints were analyzed individually to look for patterns and causes.After evaluation of the complaints, the following defects are the most common: hot plug, sig error followed by head error, error message due to shaking / error message due to sensitivity, connection problems and hardware-error.The increase in complaints with the error "head error" is due to a user / application error.The actions have been addressed in the past to prevent application errors.Furthermore, the instructions for use of the rotaflow system see rotaflow system user manual, (b)(4), contain detailed descriptions to prevent an "error head".In addition, the instruction manual describes how the user has to react in case of an error message so that the application can be continued if possible.Since there are several causes of errors that result in an error head (error head) message, no final root cause could be determined.It is noticeable that the evaluated error is largely due to a user error.Warning in the ifu regarding "hotplug" and label with the "hotplug" warning on the rfc have already been implemented in the past.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key8730375
MDR Text Key149082092
Report Number8010762-2019-00189
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW DRIVE UNIT , BLUE
Device Catalogue Number701022161
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/10/2019
Initial Date FDA Received06/25/2019
Supplement Dates Manufacturer Received09/26/2019
Supplement Dates FDA Received10/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-