The device involved in the reported incident is not available for evaluation.As such, a failure analysis could not be conducted.A review of the lot records was conducted, and did not find any indication of problems that could have caused or contributed to the complaint.As the implant was not returned for investigation, a possible root cause could not be found.The reported complaint was not confirmed.
|
A distributor reported that the customer was unhappy about the lack of backup sizes for an spip-520-0-ww silicone pip in a case they had done.On (b)(6) 2018, the patient came in for hand surgery due to joint arthritis: right long finger pip joint and right ring pip joint arthroplasty.During surgery, the surgeon decided to put in a size #0 for one finger, and then needed the same size on a second finger.The implant bank that was shipped to the facility only contained one size #0.The surgeon decided not to use a different size and instead closed the patient and would revisit to do the other finger.Now the patient will need to go through another surgery and post op therapy.There was never a plan to have to come back for further procedures on this patient.
|