DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER Back to Search Results
Catalog Number 228143
Device Problems Break (1069); Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/28/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported by the affiliate via email that the tip of the meniscal applier omnispan was defective.When the needle was positioned on the meniscus to do the suture, the tip of the gun was blocked, broken and the needle was not deploying.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was not returned after multiple attempts for device return, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.No nonconformances were identified for this part number, lot number combination per qlik query executed on 6/6/2019.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL DEPLOYMENT GUN
Type of Device
ORTHOPAEDIC CERCLAGE APPLIER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8750358
MDR Text Key150140950
Report Number1221934-2019-57499
Device Sequence Number1
Product Code GEF
UDI-Device Identifier10886705010059
UDI-Public10886705010059
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Catalogue Number228143
Device Lot Number2L82112
Date Manufacturer Received07/24/2019
Patient Sequence Number1
-
-