• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER Back to Search Results
Catalog Number GSX0025A
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 06/04/2019
Event Type  malfunction  
Manufacturer Narrative
A review of the manufacturing records verified the lot was processed normally and all pre-release specifications were met.
 
Event Description
It was reported the physician selected a 25 mm gore® cardioform septal occluder to close an atrial septal defect balloon sized to 10 mm.The device was deployed and locked without issue.As the physician retracted the retrieval cord he noted tension and the cord broke.There was cord still available on the other end and a hemostat was used to try and remove it but the cord broke again, leaving a small piece attached to the right atrial eyelet.The physician attempted to snare the remaining piece but was unsuccessful.The device appeared stable and the physician opted to leave the device implanted with the piece of cord attached.The patient was doing well following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
TRANSCATHETER, SEPTAL OCCLUDER
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
marci stewart
1500 n. 4th street
flagstaff, AZ 
9285263030
MDR Report Key8751498
MDR Text Key150279487
Report Number2017233-2019-00493
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00733132631025
UDI-Public00733132631025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/04/2020
Device Catalogue NumberGSX0025A
Device Lot Number19963514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age77 YR
Patient Weight57
-
-