The reported event was unconfirmed.Evaluation of the returned sample, submitted by futurematrix interventional, stated that when the dilator was re-inserted into the sheath to check for any resistance, resistance was felt but the dilator did transition into the sheath.The dimensional evaluation indicated that both the sheath and the dilator met specifications.The report also stated that all dhr's were reviewed, parts meet specification requirements at initial qc check and no errors were found.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following."directions for use.1.Activate the hydrophilic coating by placing the dilator and sheath components into saline or sterile water.Place an 0.035" (0.889mm) or 0.038" (0.965mm) guidewire into the ureter using standard endourology techniques, 2.Ensure the dilator lock is securely engaged with sheath hub prior to insertion, 3.Insert the guidewire into the tapered end of the dilator/sheath assembly and gradually advance the assembly into the ureter.Note: placement of the assembly can be verified using fluoroscopy or radiographic means.4.While maintaining sheath position, disengage the dilator lock from the sheath hub to gently remove the dilator.Do not advance sheath without the dilator in place.Note: suture holes are provided on sheath hub for securing externally, if desired, 5.An endoscope and/or related instruments can now be used through the ureteral sheath as needed, 6.If desired, irrigation can be applied using the luer connector on the dilator, 7.Upon completion of the access procedure, gently withdraw the device, 8.Discard the device in accordance with hospital procedures and with applicable laws and regulations.".
|