Catalog Number 8065990710 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/20/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.(b)(4).
|
|
Event Description
|
A facility representative reported during measurement testing prior to refractive treatment, the measurements were not correct.The measurements were not used during surgery and there was no patient impact.
|
|
Manufacturer Narrative
|
During technical review the field service engineer (fse) performed the calibration with model eye and found all measurement values in range.Measurement was also checked on human eyes and found to be normal.No technical root cause was identified as the product was found to be within specifications.The root cause could not be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|