CONCORD MANUFACTURING 2008K@HOME HEMODIALYSIS SYSTEM; HEMODIALYSIS SYSTEM FOR HOME USE Back to Search Results
Catalog Number 190828
Device Problems Mechanical Problem (1384); Reflux within Device (1522)
Patient Problem No Patient Involvement (2645)
Event Date 06/21/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A user facility biomed technician reported to technical support that a 2008k machine experienced a bag of saline filling up.The saline bag backfill occurred during recirculation.The patient was not connected during recirculation.The biomed reported that there were no quick disconnects on the drain line and that the drain line length and height were within specification.The machine was returned to service without a recurrence of the reported issue.An air separator conductivity cable was changed.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K@HOME HEMODIALYSIS SYSTEM
Type of Device
HEMODIALYSIS SYSTEM FOR HOME USE
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8779788
MDR Text Key150712943
Report Number2937457-2019-02103
Device Sequence Number1
Product Code ONW
UDI-Device Identifier00840861100941
UDI-Public00840861100941
Combination Product (y/n)N
PMA/PMN Number
K070049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 07/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190828
Was Device Available for Evaluation? No
Device Age MO
Date Manufacturer Received07/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-